Construction of a novel highly selective potentiometric perchlorate sensor based on neocuproine-Cu(II) complex formed in situ during the conditioning period


Onder A., Topcu C., ÇOLDUR F.

Chemija, cilt.29, sa.1, ss.57-66, 2018 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 29 Sayı: 1
  • Basım Tarihi: 2018
  • Dergi Adı: Chemija
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus
  • Sayfa Sayıları: ss.57-66
  • Anahtar Kelimeler: Electrode, Neocuproine, Perchlorate, Perchlorate determination, Potentiometry
  • Erzincan Binali Yıldırım Üniversitesi Adresli: Evet

Özet

© Lietuvos mokslu akademija, 2018.A potentiometric perchlorate-selective electrode based on a phenanthroline derivative, neocuproine (2,9-dimethyl-1,10-phenanthroline), was developed. An electroactive substance, a neocuproine complex of Cu2+, providing a selective response to perchlorate was formed in situ in the electrode membrane structure during the conditioning process. Membrane optimization studies showed that the most suitable membrane composition was 1.0% (w/w) neocuproine, 69.0% (w/w) o-nitrophenyloctylether (o-NPOE) and 30.0% (w/w) poly(vinyl chloride) (PVC) in terms of potentiometric performance characteristics. For the proposed electrode, the linear range, slope, detection limit, pH working range and response time were determined as 2.5 × 10-7-1.0 × 10-1 M, -53.01 mV/decade, 1.0 × 10-7 M, 4.0-12.5 and ≈3 s, respectively. The proposed sensor was successfully used in the determination of perchlorate contents of some perchlorate spiked tap, river and dam water samples.